EasyManuals Logo

Yuwell YE660D User Manual

Yuwell YE660D
115 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #23 background imageLoading...
Page #23 background image
Electromagnetic compatibility information
Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture' s declaration- electromagnetic emission
The YE660D is intended for use in the electromagnetic environment specified
below. The customer or the user of the YE660D should assure that it is used in
such and environment.
Immunity test
IEC 60601 test level
Compliance level
Conducted RF
IEC 61000-4-6
3 Vrms
0,15 MHz-80 MHz
6 V rms in ISM and
amateur radio bands
between 0,15 MHz and
80 MHz
80 % Am at 1 kHz
3 Vrms
0,15 MHz-80 MHz
6 V rms in ISM and
amateur radio bands
between 0,15 MHz and
80 MHz
80 % Am at 1 kHz
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to
2.7 GHz
10 V/m
80 MHz to
2.7 GHz
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people
a) Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the YE660D is used exceeds the
applicable RF compliance level above, the YE660D should be observed to verify
normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the YE660D.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3 V/m
Electromagnetic compatibility information
Table 4 - Test specifications for ENCLOSURE PORT IMMUNITY to RF
wireless communications equipment
Test
frequency
(MHz)
Band
a)
(MHz)
Service
a)
(MHz)
Modulation
b)
Maximu
m power
(MHz)
Distanc
e (m)
IMMUNIT
Y TEST
ELVEL
(V/m)
385
380-390
TETRA 400
Pulse
modulation
b)
18
Hz
1,8
0,3
27
450
430-470
GMRS 460, FRS
460
FM±5 kHz
deviation
1 kHz sine
2
0,3
28
710
704-787
LTE Band 13, 17
Pulse
modulation
b)
21
7 Hz
0,2
0,3
9
745
780
810
800-960
GSM
800/900,TETRA
800,iDEN
820,CDMA
850,LTE Band 5
Pulse
modulation
b)
18
Hz
2
0,3
28
870
930
1720
1700-
1990
GSM 1800;
TETRA 1900;
GSM 1900;
DECT;LTE Band
1,3, 4,25; UMTS
Pulse
modulation
b)
21
7 Hz
2
0,3
28
1845
1970
2450
2400-
2570
Bluetooth,WLAN,
802.11 b/g/nRFID
2450,LTE Band 7
Pulse
modulationb)21
7 Hz
2
0,3
28
5240
5100-
5800
WLAN802.11a/n
Pulse
modulationb)21
7 Hz
0,2
0,3
9
5500
5785
NOTE If necesary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting
antenna and theME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is
permitted by IEC61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it
does not

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Yuwell YE660D and is the answer not in the manual?

Yuwell YE660D Specifications

General IconGeneral
BrandYuwell
ModelYE660D
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals