Electromagnetic compatibility information
Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture' s declaration- electromagnetic emission
The YE660D is intended for use in the electromagnetic environment specified
below. The customer or the user of the YE660D should assure that it is used in
such and environment.
Conducted RF
IEC 61000-4-6
3 Vrms
0,15 MHz-80 MHz
6 V rms in ISM and
amateur radio bands
between 0,15 MHz and
80 MHz
80 % Am at 1 kHz
3 Vrms
0,15 MHz-80 MHz
6 V rms in ISM and
amateur radio bands
between 0,15 MHz and
80 MHz
80 % Am at 1 kHz
Radiated RF
IEC 61000-4-3
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people
a) Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the YE660D is used exceeds the
applicable RF compliance level above, the YE660D should be observed to verify
normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the YE660D.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3 V/m
Electromagnetic compatibility information
Table 4 - Test specifications for ENCLOSURE PORT IMMUNITY to RF
wireless communications equipment
IMMUNIT
Y TEST
ELVEL
(V/m)
FM±5 kHz
deviation
1 kHz sine
GSM
800/900,TETRA
800,iDEN
820,CDMA
850,LTE Band 5
GSM 1800;
TETRA 1900;
GSM 1900;
DECT;LTE Band
1,3, 4,25; UMTS
Bluetooth,WLAN,
802.11 b/g/nRFID
2450,LTE Band 7
NOTE If necesary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting
antenna and theME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is
permitted by IEC61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.、
c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it
does not