HFA3 Instructions for Use 2660021166131 Rev. A 2018-11
Go to Contents Data Transfer
9-1
(9) Data Transfer
This chapter describes how to move tests from an older Humphrey Field Analyzer model (either the
HFA II or an HFA II-
i
) to an HFA3 in local database mode. Transfer data by using the instrument’s
network connection, a USB device, or by connecting a serial cable to the Data Transfer ports of two
HFA instruments. Serial connections can be: HW COM port, COM-USB adapter or COM-Network
adapter. The HFA3 is able to combine HFA II, HFA II-
i
and HFA3 data to create reports.
To transfer data from an HFA II or HFA II-
i
to the FORUM server see the
Humphrey
®
HFA
™
Connect
Software Installation Instructions
. Data on the server can then be accessed by an HFA3 in FORUM
connected mode.
Set Up a Serial Transfer from an HFA II or HFA II-
i
To set up a data transfer from an HFA II or HFA II-
i
:
1. Go to Se
ttings > Specific Settings > HFA Data Transfer.
2. Follow the detailed instructions on the screen to set up a serial
tr
ansfer between an HFA II or HFA II-
i
and an HFA3.
3. Select Next once the two
instruments are connected.
4. In the next screen, choose On
to add a patient ID number during
data transfer. This feature is required for data import. Note: A unique
29-character Patient ID will be automatically created for the patient
from the patient’s name and date of birth.
5. Name splitting is a mandatory feature automatically set to On.
Not
e: HFA II and HFA II-
i
(software versions before 5.0) used one
name field for patient records. To generate a report using patient
data on the HFA3, you must change the patient record to include a
separate first and last name. See “Name Splitting,” on page 9-2.
6. Select Next
to proceed to the next screen.