Error Messages / Functional test / Legal Notices / Disposal
錯誤訊息 / 功能測試 / 法律聲明 / 棄置
錯誤代碼列表
224/1通訊錯誤
402/2超聲波治療頭過熱 407/2 PLL故障
406/2超聲波頭未接觸 410/2輸出電壓異常
某些情況下,重新啟動電源可解決問題。只能由專業人員維修。另提供服務手冊。
功能測試
用耦合凝膠覆蓋超聲波頭,設定強度,開始治療,耦合值必須達90 %以上。
Error Code list
224/1 Communication Error
402/2 Ultrasound applicator has overheated 407/2 PLL fault
406/2 Ultrasound head not connected 410/2 Output Voltage abnormal
In some cases switching the device on an off may solve the problem. Service and repair only by trained
staff. Service manual available.
Functional test
Cover ultrasound head with coupling gel, set intensity, start therapy. Coupling display must indicate
over 90 %.
如有疑問或設備出現故障,請聯繫分銷商。
For questions or in case of equipment malfunctions, please contact the distributor.