57
Performance Characteristics
56
Regression statistics
Model PG110 (2in1. MICRO
Blood Glucose
Monitoring System)
Test Method Electrochemical
sensor
Calibration Method Plasma
Sample Whole blood,
capillary
Sample Size 0.5 μL
Test Time 5 seconds
Memory No limit
Result Range 10~600mg/dL
(0.6~33.3mmol/L)
Hematocrit 20-60%
Operating
Temperature
10~40°C (50~104°F)
Operating
Relative Humidity
10-90%
Altitude Up to 3,048 meters
(10,000 feets)
Power Source From the smart
phone
Size 41.5 mm x 33.6 mm
x 9.0 mm
Weight 5.9 g (0.21 oz)
Within run precision
(100 venous blood tests per glucose level)
Total precision
(100 control solution tests per glucose level)
Slope 0.987
Intercept 0.469 mg/dL
R
2
0.989
No. of samples 100
Range tested 35.6~422.4 mg/dL
Mean glucose
Standard
deviation
(mg/dL)
Coecient
of variation
(%)
41 mg/dL (2.3 mmol/L) 1.2 2.8
99 mg/dL (5.5 mmol/L) 2.7 2.7
131 mg/dL (7.3 mmol/L) 2.8 2.1
231 mg/dL (12.8 mmol/L) 3.7 1.6
373 mg/dL (20.7 mmol/L) 8.0 2.1
Mean glucose
Standard
deviation
(mg/dL)
Coecient
of variation
(%)
43 mg/dL (2.4 mmol/L) 1.0 2.2
104 mg/dL (5.8 mmol/L) 2.4 2.3
297 mg/dL (16.5 mmol/L) 8.9 3.0
Specifications