EasyManua.ls Logo

A.W. Sperry Instruments DSA-2002P - Page 8

A.W. Sperry Instruments DSA-2002P
24 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
―6―
●Operating System Dual Integration
●Display Liquid crystal display with a maximum
count of 4000
●Low Battery Warning "BATT" symbol is displayed on the digital
display.
●Overrange Indication "OL" is displayed where input exceeds
the upper limit of a range
●Response Time Approx. 2 seconds
●Sleep function Automatically powered down in about 10
minutes after the last switch operation
●Data Hold Available in all ranges provided the peak
measurement mode is deactivated.
●Location for use Indoor use, Altitude up to 2000m
●Storage Temperature ー20〜60℃, relative humidity up to 85%
and Humidity without condensation
●Operating Temperature 0〜40℃, relative humidity up to 85%
and Humidity without condensation
●Conductor Size Approx. 54.5㎜ diameter max.
●Overload Protection 2400A AC for 10sec
1200V AC/DC for 10sec
600V AC for 10sec
●Withstand Voltage 5550V AC for 1 minute between electrical
circuit and housing cases or metal parts
of jaws
●Insulation Resistance 10MΩ or greater at 1000V between
electrical circuit and housing cases or
metal parts of jaws
●Safety Standard IEC 61010-1:Measurement CAT.Ⅲ
600VAC,DC / CAT.Ⅱ750VAC, 1000VDC,
pollution degree2.
●Dimensions 247(L)×105(W)×49(D)mm
●Weight Approx. 470g(battery included)
●Power Source Two R6P(DC1.5V) batteries or equivalent
●Current Consumption Approx. 5mA max. (Approx. 20μA in the
sleep mode)
●Accessories Test leads M-7107
Two R6P batteries
Instruction manual
Recorder Output Plug M-8201
Carrying case M-9094
●Optional Accessories Output Probe M-7014
Recorder M-5100A, etc.