D
Danielle FarrellAug 19, 2025
What to do if the Aalborg XFM LCD display is blank?
- DDon SnyderAug 19, 2025
If the Status LED indicator and LCD Display on your Aalborg Measuring Instruments remain blank after powering on, and there's no response from analog outputs, it could be due to a bad or reversed power supply. First, measure the voltage on pins 7 and 5 of the 15 pin D-connector; if the voltage is outside the specified range, replace the power supply. Also, check the polarity; if it's reversed, correct the connection. If the problem persists, the PC board may be defective, requiring you to return the XFM to the factory for repair.