EasyManuals Logo

Aalborg XFM Series User Manual

Aalborg XFM Series
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #38 background imageLoading...
Page #38 background image
35
UART Error Codes:
1 - Not Supported Command or Back Door is not enabled.
2 - Wrong # of Arguments.
3 - Address is Out of Range (MR or MW commands).
4 - Wrong # of the characters in the Argument.
5 - Attempt to Alter Write Protected Area in the EEPROM.
6 - Proper Command or Argument is not found.
7 - Wrong value of the Argument.
8 - Reserved.
9 - Manufacture specific info EE KEY (wrong key or key is disabled).
COMMAND
NAME
DESCRIPTION
No.
COMMAND SYNTAX
Command Argument 1
Argument 2
Argument 3 Argument 4 Response
Maintenance
Timer
Hours since last time unit
was calibrated.
11 C
R (read timer)
<Value> (in Hours)
C(set timer to zero)
CC
Full Scale
Returns the full scale rated flow in
L/min. (Note: This term is not
multiplied by the current K factor)
12
E
<Value>
(in L/min)
LCD Back
Light
LCD Back Light control
(0-100.0%)
0 - off
100 - Maximum Intensity
13
B 0 to 100%
B:<Counts>
where:
Counts (0 – 4095)
B:<Value>
where:
Value (0 – 100.0)
No Argument
<Current settings>
Read
EEPROM
Memory
Reads the value in the
specified memory location.
14
MR 0000 to999
(Table Index)
<value>
Write
EEPROM
Memory
Writes the specified value to the
specified memory location. Use
Carefully, can cause unit to
malfunction. (Note: Some addresses
are write protected!)
15 MW 0000 to 999
(Table Index)
Value MW,XXX,<Value>
where:
XXX=Table Index

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Aalborg XFM Series and is the answer not in the manual?

Aalborg XFM Series Specifications

General IconGeneral
BrandAalborg
ModelXFM Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals