UART Error Codes:
1 - Not Supported Command or Back Door is not enabled.
2 - Wrong # of Arguments.
3 - Address is Out of Range (MR or MW commands).
4 - Wrong # of the characters in the Argument.
5 - Attempt to Alter Write Protected Area in the EEPROM.
6 - Proper Command or Argument is not found.
7 - Wrong value of the Argument.
8 - Reserved.
9 - Manufacture specific info EE KEY (wrong key or key is disabled).
COMMAND
NAME
DESCRIPTION
No.
COMMAND SYNTAX
Command Argument 1
Argument 2
Argument 3 Argument 4 Response
Maintenance
Timer
Hours since last time unit
was calibrated.
11 C
R (read timer)
<Value> (in Hours)
C(set timer to zero)
CC
Full Scale
Returns the full scale rated flow in
L/min. (Note: This term is not
multiplied by the current K factor)
12
E
<Value>
(in L/min)
LCD Back
Light
LCD Back Light control
(0-100.0%)
0 - off
100 - Maximum Intensity
13
B 0 to 100%
B:<Counts>
where:
Counts (0 – 4095)
B:<Value>
where:
Value (0 – 100.0)
No Argument
<Current settings>
Read
EEPROM
Memory
Reads the value in the
specified memory location.
14
MR 0000 to999
(Table Index)
<value>
Write
EEPROM
Memory
Writes the specified value to the
specified memory location. Use
Carefully, can cause unit to
malfunction. (Note: Some addresses
are write protected!)
15 MW 0000 to 999
(Table Index)
Value MW,XXX,<Value>
where:
XXX=Table Index