BM series 73
15.3.4. Command Examples
This example uses the "erCd on" of "5if" so that the <AK> (06h) code is output.
ON command (turning on the balance)
Computer Time Balance
ON command
Standby mode
Command received
All segments are displayed
Command completed
Zero display
R command (Re-zeroing the display)
Computer Time Balance
R command
Weighing mode
Command received
In process
Command completed
Zero display
CAL command (Calibration with internal mass)
Computer Time Balance
CAL command
Zero display
Command received
Command completed
Zero display
All segments