62 BM series
Calibration test report using the internal mass
Note Calibration test does not perform adjustment.
If the data memory function is used (except data 0), the "Title block" and "End
block" cannot be output.
Key operation
1 Press and hold the CAL key to display CC in and release the key.
2 CC is displayed and the balance is checked automatically.
3 The zero point is measured and the weight value is displayed for a few seconds.
4 The internal mass is weighed and the weight value is displayed for a few seconds.
5 If GLP output is used, glp is displayed and the calibration test report is output.
6 The balance returns to the weighing mode.
Command
This calibration test can be performed with the TST command.
Function setting "info 1" Function setting "info 2"
AD-8121 format
General format using the balance built-in clock data
Space, ASCII 20h
<TERM>
Terminator, CR , LF or CR
CR Carriage return, ASCII 0Dh
LF Line feed, ASCII 0Ah
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration test type
Zero point value
Target weight value
Target weight
Signature
-----------A-&-D
MODEL-----BM-300
S/N-----01234567
ID------LAB-0123
DATE--2011/02/20
TIME----12:34:56
CAL.TEST(INT.)
ACTUAL
-------0.0000--g
----+200.0002--g
TARGET
----+200.0000--g
SIGNATURE
---------------