B
bmasonSep 10, 2025
What does excess characters error mean on A&D GX-200?
- JJohn MaysSep 10, 2025
An excess characters error on A&D Scales means the balance received excessive characters in a command. Confirm the command to resolve this issue.
What does excess characters error mean on A&D GX-200?
An excess characters error on A&D Scales means the balance received excessive characters in a command. Confirm the command to resolve this issue.
How to fix stability error on A&D GX-200?
A stability error on A&D Scales typically arises from environmental factors. To resolve this, prevent any vibration, drafts, temperature changes, static electricity, and magnetic fields around the scales. You can also adapt the balance to the environment.
What causes calibration weight error on A&D GX-200?
A calibration weight error on A&D Scales occurs when the calibration weight is either too heavy or too light. Confirm the correct calibration weight value. Press the CAL key to return to the weighing mode.
Why am I getting a communications error on my A&D GX-200?
A communications error on A&D Scales indicates a protocol error during communication. To fix this, confirm the correct format, baud rate, and parity settings.
What to do if A&D GX-200 show weighing pan error?
If your A&D Scales display a weighing pan error, ensure that the weighing pan is correctly installed. After confirming its installation, calibrate the balance.
What does timeout error mean on A&D GX-200 Scales?
A timeout error on A&D Scales means the balance didn't receive the next character of a command within one second. Confirm the communication to resolve this.
How to resolve undefined command error on A&D Scales?
To resolve an undefined command error on A&D Scales, confirm the command you are trying to execute.
What causes automatic response adjustment zero error on A&D Scales?
The automatic response adjustment zero error on A&D Scales indicates that there is something on the pan. Clear the pan and press the CAL key to return to the weighing mode.
How to fix clock battery error on A&D GX-200?
If your A&D Scales display a clock battery error, it means the clock backup battery has been depleted. Press any key and set the time and date to resolve this.
What does parameter setting error mean on A&D Scales?
A parameter setting error on A&D Scales means the received data exceeds the range that the balance can accept. To resolve this, confirm the parameter range of the command.
Capacity | 210 g |
---|---|
Readability | 0.001 g |
Repeatability (Standard Deviation) | 0.001 g |
Linearity | ±0.002 g |
Stabilization Time | Approx. 1.5 seconds |
Calibration | External calibration |
Interface | RS-232C |
Operating Temperature | 5°C to 40°C |
Display | LCD |
Power Source | AC adapter |
Units | g, oz, ozt, ct, dwt, GN, tl |