Running a Blank
A blank must be run every 12 hours. Each blank measurement remains
valid for 12 hours of continuous operation, aer which the analyzer displays
BLANK needed and a new blank is required. In addition, a new blank must
be run each time the analyzer has been powered on and off again.
If the analyzer is in standby mode, touch the measure icon.
Touch the Run button.
If a blank is necessary, the “Blank needed” screen appears.
Touch OK to confirm and run the blank.
Check the results to see whether the blank falls within the acceptable
ranges shown in the following table. If so, press Accept.
Acceptable blanks will have
the following readouts:
WBC 0.0-0.5
RBC 0.0-0.05
PLT 0-26
HGB 0-10
.
2