Changing a component 
The (maximum four) components, which are displayed in the component list 
(see page 165), can be changed. Either the parameters of a component can 
be changed or one component can be replaced by another.   
Menu path 
MENU → Configure → Component-specific → Active Component 
Select component and confirm with ENTER. 
Change parameter of a component 
NOTES 
In the case of a system component (Type S) only the parameter "Unit" can be 
changed; in the case of a user component (Type B) all the parameters can be 
changed with the exception of the parameter "Type".   
The parameters of the active component (Type SA or BA) and the compo-
nent used for calibration (Type SK or BK) cannot be changed. 
1  Press softkey COMP. CONFIG. The "Configure Component" menu is 
displayed with the component list. 
2  In the component list select the component with parameters which are 
to be changed. 
 
 
 
3  Press softkey CHANGE COMP. Enter password (level 3), as applicable. The 
parameters of the component are displayed.