AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  327
 
 
 
No. Text  S A 
F Reaction/Remark 
358  Lamp intensity above or below 
middle of permissible range. 
        x      Lamp intensity has fallen to 10% of Init value. Check the 
intensity values in the Diagnostics /Test → Module 
specific
 → Lamp Intensity menu. If all four values 
have fallen by about the same amount compared to the 
Init values, the cause is reduced lamp intensity. The lamp 
should be soon replaced (see page 307). If only the two 
measurement detector values have dropped, the cause 
is probably a contaminated sample cell. Clean the sam-
ple cell (see page 294) or exchange it. 
359  Lamp intensity above or below 
permissible range. 
x  x          Lamp intensity has fallen to 5% of Init value. Check the 
intensity values in the Diagnostics /Test → Module 
specific 
→ Lamp Intensity menu. If all four values 
have fallen by about the same amount compared to the 
Init values, the cause is reduced lamp intensity. Replace 
the lamp (see page 307) and perform an amplification 
optimization (see page 310).   
360  Filter wheel 1 cannot be initial-
ized. 
x  x      Notify Service. 
361  Filter wheel 2 cannot be initial-
ized. 
x  x      Notify Service. 
362  The calibration filter wheel 
cannot be initialized. 
x  x      Notify Service. 
363  The Limas analyzer board can-
not be initialized. 
x  x      Notify Service. 
364  A new lamp has been installed. 
The amplifiers settings are 
optimized. 
        For information 
365  The Limas amplifiers settings 
are optimized. 
        For information 
LS25             
366  The LS25 analyzer has an overall 
error. 
x  x       
367  The LS25 analyzer has a 
maintenance request. 
x    x     
368  The LS25 analyzer starts meas-
urement.  
x  x       
369  LS25 analyzer detector error #  x  x             
370  Beam transmission too low          x      The beam transmission has fallen below the limit value 
for reliable measurements. This indicates improper 
alignment of the optical units or contamination of the 
optical windows. Clean optical windows and check 
alignment. 
371  The temperature input signal 
exceeds the permissible range. 
        x      The gas temperature is outside the permissible range 
for temperature equalization. The accuracy of the meas-
ured gas concentration could have been reduced. If the 
supplied values are correct, the device is operating out-
side the specifications. 
372  The pressure input signal ex-
ceeds the permissible range.   
    x    The gas pressure is outside the permissible pressure 
equalization range. The accuracy of the measured gas 
concentration could have been reduced. If the supplied 
values are correct, the device is operating outside the 
specifications.