Receiver Size Height: 2.5 in. (6.3 cm). Width: 3.24 in. (8.2 cm).
Depth: 0.88 in. (2.2 cm).
Receiver Weight 3.5 oz. (99.2 grams) – including batteries.
Receiver Battery Life Up to 60 days.
Automatic Shuto Built-in blood glucose meter: 2 minutes aer last user
action. Receiver: 12 seconds aer last user action.
Receiver Memory •60daysofnormaluseincludingcontinuousglucose
readings (stored every 10 minutes) and daily blood
glucose readings.
•Date/timewillberememberedfor 5 minutes aer receiver
battery removal.
Calibration Plasma equivalent.
Calibration Time •1stcalibration:Mustbeperformedatapproximately
10 hours aer a new sensor has been inserted. e
rst calibration can be performed aer the 10 hours.
Continuous glucose readings will NOT be reported until
the 1st calibration is performed successfully.
•2ndcalibration:Mustbeperformedbetween2and4hours
aer the 1st calibration or continuous glucose will not
be reported. e 2nd calibration can be performed aer
4 hours; continuous glucose reporting will resume aer
completing the 2nd calibration successfully.
•3rdcalibration:Mustbeperformedbetween12and20
hours aer the 2nd calibration or continuous glucose will
not be reported. e 3rd calibration can be performed
aer 20 hours; continuous glucose reporting will resume
aer completing the 3rd calibration successfully.
•4thcalibration:Mustbeperformedbetween48and56
hours aer the third calibration or continuous glucose will
not be reported. e fourth calibration can be performed
aer 56 hours; continuous glucose reporting will resume
aer completing the 4th calibration successfully.
Blood Glucose Test Time
(for BG mode)
Average of 7 seconds.
Blood Glucose Assay Method
(for BG mode)
Coulometric electrochemical sensor.
Blood Sample Type (for BG mode) Whole blood, capillary.
Hematocrit (for BG mode) 15% to 65%.
18 Appendix B: Specications (con’t)
Customer Care: 1-866-597-5520
152 Section 18
Appendix B: Specications