ABELL A-82 UHF Service Manual
16 41
ADJUSTMENT
Required test Equipment
1
Stability power supply
1)The supply voltage can be changed between 5V and 18V, and the current is 3A or more.
2)The standard voltage is 7.5V
2
DC Ammeter
1). Class 1 ammeter (17ranges and other features).
2). The full scare can be set to either 300mA or 3A.
3). A cable of less internal loss must to be used.
3
Frequency counter (F. Counter)
1)Frequency of 1GHz or so can be measured.
2)The sensitive can be changed to 500MHz or blow, and measurements are highly stable and accurate.
(about 0.2ppm).
4
Power Meter
1). Measurable frequencyUp to 500MHz
2). Impedance50Ω, instability.
3). Measuring rangefull scare of 10W or so.
4). A standard cable (5D2W 1m) must be used.
5
RF Voltmeter(RF V.M)
1)Frequency rangeup to 500MHz or so.
6
Linear Detector
1). Frequency rangeup to 500MHz or so.
2). Characteristics are flat, CN is 60dB or more.
7
Digital Voltmeter
1). Voltage rangeFS=18V or so.
2). Input resistance: 1MΩ or more.
8
Oscillograph
1). Measuring rangeDC to 30MHZ.
2). Provide highly accurate measurements for5 to 25MHz.
9
AF Voltmeter (AF V.M)
1). Measuring range50Hz to 1MHz
2). Maximum sensitivity1mV or more.