In-service plan for blood glucose testing
ACCU-CHEK Inform II Calibration (Coding) Procedure:
1. Turn the meter on.
2. Enter your operator ID via the barcode scanner or manually according to facility policy.
3. Touch to conrm and display the Main Menu.
4. Touch to open the Main Menu 2 screen.
5. Touch Strip Lots to open the related menu.
6. Touch Add to upload new test strip lot information from a new code key. The Add Strip Lot
screen opens.
7. Insert the new code key in the opening of the code key reader. An LED ashes and signals
that the reader is ready to upload data.
8. Place the code key reader on a level surface, such as a bench or table. Hold the meter
4-6 inches above the code key reader so that a connection can be made between the two
infrared windows.
9. Touch on the meter to begin uploading data.
10. Information about the expiration date and parameters for control solutions is subsequently
displayed.
11. Touch to store the data for this lot number in the meter without changes or touch
to modify the data for this lot number before storing it in the meter.
12. Once you have nished updating the test strip information, touch to conrm that you
want this lot number to be the lot number currently in use.
or
13. Continue entering additional lot numbers.
or
14. Touch to return to the Main Menu.

Other manuals for Accu-Chek Inform II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Inform II and is the answer not in the manual?

Accu-Chek Inform II Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Memory Capacity1000 test results
DisplayLCD
Measurement Range10 to 600 mg/dL
ConnectivityInfrared
Sample TypeCapillary whole blood
Operating Temperature16 to 32°C (61 to 90°F)
Relative Humidity10 to 90% non-condensing