PHSA Laboratories CW Site - Point of Care
Title: CWPC_GLU_0140 Accu-Chek Inform II Training Guide
Medical Approval: Dr Benjamin Jung Medical Approval Date: 27 Dec 2016
Version: 1.2 Implementation Date: 7/2/2019 11:55:59 PM
Folder Name: CW\Point of Care\Whole Blood Glucose
This is a controlled document for CW use only. Any printed copies are uncontrolled unless specified. Please refer to Lab QMS
Page 9 of 12
Can simulate with a Level 5 or 6 Proficiency sample during training.
Meter is ready for Patient sample.
Follow reference guides for the capillary puncture.
Select puncture site
Choose appropriate capillary puncture device
Puncture site is clean. Clean site with soap & water if required.
Disinfect with 2% chlorhexidine in 70% isopropyl alcohol
Perform skin puncture.
Remove first drop of blood. (removes any contaminants or residual
disinfectant solution which could dilute the glucose value).
Apply 2
nd
drop to the FRONT EDGE of the test strip.
Result appears after 5 seconds
Numerical Glucose value will display if within 0.6-33.3 mmol/L
@ PHSA there are two different configurations programmed:
Configuration 1 Children’s Hospital
Women’s Hospital
Neonate ranges Sunny Hill
Displayed Canuck Place
Configuration 2 BC Cancer Agency – all wards & clinics
Adult ranges
Displayed

Other manuals for Accu-Chek Inform II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Inform II and is the answer not in the manual?

Accu-Chek Inform II Specifications

General IconGeneral
Sample Size0.6 µL
Test Time5 seconds
Memory Capacity1000 test results
DisplayLCD
Measurement Range10 to 600 mg/dL
ConnectivityInfrared
Sample TypeCapillary whole blood
Operating Temperature16 to 32°C (61 to 90°F)
Relative Humidity10 to 90% non-condensing