Specifications
Model VIC-303 VIC-123 VIC-4MG VIC-2MG VIC-612 VIC-412 VIC-212
Weighing capacity g 300 120 410 210 610 410 210
Readability g 0.001 0.001 0.005 0.005 0.01 0.01 0.01
Tare range (subtractive) g 300 120 410 210 610 410 210
Linearity ≤±g 0.004 0.003 0.01 0.01 0.03 0.03 0.035
Operating temperature range 10°C to 30°C (273°K to 303°K; 50°F to 86°F)
Stabilization time (average) s 2.5 2.5 2.5 2.5 2 2 2
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels; display update: 0.1–0.8 (depends on filter level selected)
Calibration weight g 200 (F1) 100 (F1) 200 (F1) 200 (F2) 200 (F2) 200 (F2) 200 (M1)
Net weight, approx. kg 1.3 1.2 1.3 1.3 1.35 1.35 1.2
Pan size mm 97 d 97 d 97 d 97 d 142x130 142x130 97 d
Power source/voltage/frequency AC adapter, 230 V or 115 V, +15% to – 20%, 48–60 Hz
Power consumption (average) W 1 1 1 1 0.75 0.75 0.75
Hours of operation w/ 9V battery:
– Alkaline (approx.) h – – – – 11 11 14
– Rechargeable, fully ch.,
(NiMH), avg. h – – – – 2.5 2.5 4
Model
VIC-5101 VIC-3101 VIC-1501 VIC-711 VIC-511 VIC-10KG VIC-6KG VIC-4KG
Weighing capacity g 5100 3100 1500 710 510 10100 6100 4100
Readability g 0.1 0.1 0.1 0.1 0.1 1 1 1
Tare range (subtractive) g 5100 3100 1500 710 510 10100 6100 4100
Linearity ≤±g 0.2 0.2 0.2 0.2 0.2 2 2 1
Operating temperature range 10°C to 30°C (273°K to 303°K; 50°F to 86°F)
Stabilization time (average) s 2 2 1.5 1.5 1.5 1.5 1.5 1.5
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels; display update: 0.1–0.8 (depends on filter level selected)
Calibration weight kg 5 (F2) 2 (F2) 1 (M1) 0,2 (M2) 0,2 (M2) 5 (M1) 5 (M2) 2 (M2)
Net weight, approx. kg 1.1 1.1 1.1 1.25 1.25 1.1 1.1 1.1
Pan size mm 142x130
Power source/voltage/frequency AC adapter, 230 V or 115 V, +15% to – 20%, 48–60 Hz
Power consumption (average) W 1 1 0.75 0.75 0.75 0.75 0.75 0.75
Hours of operation w/ 9V battery:
– Alkaline (approx.) h 11 11 14 14 14 14 14 14
– Rechargeable, fully ch.,
(NiMH), avg. h 2.5 2.5 4444 44