2. Error occurrence & Repair
Error Cause Repair
In case that the
P.B.F. is too low or
too high
• Measuring method.
Measuring is done under
unstable hydration state
such as just after
physical exercise, a
bath, sweating, drinking
or overeating.
Move or speak during
measurement
Electrode holders or
measuring parts are
dirty
Faulty in impedance
measuring
• Measure again in stable physique.
• If there is movement or speaking during
measurement, the weight and impedance are
not correct. Measure again correctly.
• Try to measure again after cleaning the
electrode holders with soft gauze.
• Try again after cleaning the hands and soles.
• Check if foreign materials are between
electrodes and measuring parts.
• Try again in correct posture and hold the
electrodes according to the measuring method.
Machine doesn’t
Work When ‘start’
button is pressed
When the electrode
holders are
disconnected from the
main body or contact is
faulty.
'Start' button is
defective.
make sure that the electrode holders are
connected to the main body well.
• If the same error occurs even after connecting
the electrodes well, please contact with company
or its local distributor from where device is
purchased.