8
Percentage Sampling Specifications
JW-1
300g 600 g 1500 g 3000g
Minimum 100% Weight
1 g 2 g 5 g 10 g
Minimum Percentage
0.01%
Maximum Percentage
30,000.0 %
NOTE:
Once a successful sample has been accepted by the
balance, the Piece Weight and the Percentage will continue to be used
until it is reset by another successful sample. This is true even if you
turn the scale off or unplug it.
Settings Mode
There are 4 parameters that can be set depending upon the balance’s
application. These are designated as User Parameters (UP) 1 thru 4.
The listed values are the default values. See the chart below for a
detailed explanation f each of the User Parameters.
Entering Settings Mode
1. Make sure that the balance is turned OFF
2. Press and hold down the ¿ key
3. Turn the scale ON
4. Once the display reads UP- release the ¿ key
5. Press the ¿ key
6. The display should read UP-1 0 where the 1 signifies the User
Parameter and the 0 is the value that it is currently set at
7. Use the SET key to toggle the value to its desired setting
8. Press the ¿ key to store the value and move on to the next User
Parameter
9. The display should read UP-2 0
10. Use the SET key to toggle the value to its desired setting
11. Press the ¿ key to store the value and move on to the next User
Parameter
12. The display should read UP-3 0
13. Use the SET key to toggle the value to its desired setting
14. Press the ¿ key to store the value and move on to the next User
Parameter
15. The display should read UP-4 0