3.3 Averager with simultaneous acquisition and readout (AVG & CST)
Acqiris SA220P User's Manual 43
Figure 3.7 - AVG & CST acquisition sequence.
Performance
Excepted for small number of averages, the architecture enables no trigger loss. The maximum
performance of the system depends on the number of averages and the trigger rate.
The data truncation is not supported in averaging mode with streaming.