Waveform (nominal value, irradiance, duration) Average irradiance over time
Pulsed mode
1020 mW / cm
2
754 mW / cm
2
Progressive mode
937 mW / cm
2
693 mW / cm
2
4.9.3 Cure depth
Cure depth in mm with a distance of 2 mm between the optical guide and the cure material
Rapid mode Pulsed mode Progressive mode
Z100 MP, 3M 3mm/2.5mm 2 mm/2.5mm 3mm/2.5mm
Tetric EvoCeram, Ivoclar Vivadent 2 mm/1.5 mm 1.5 mm/1.5 mm 2 mm/2 mm
Charisma, Heraeus - Kulzer 2 mm/2 mm 2 mm/1.5 mm 2.5 mm/2 mm
4.10 Switching off the medical device
Press the ON/OFF button to stop a cure cycle.
The Mini LED switches off after three minutes of inactivity. Press the button to restart.
4.11 Disconnecting the medical device
Before a long absence or when not in use, the medical device must be cleaned, its battery must be removed and the
charging base must be disconnected from the mains power.