10 - ELECTROMAGNETIC COMPATIBILITY
Warning: The charger power cord must be kept apart from those of any nearby devices.
M
INI L.E.D. requires special precautions to be taken with regard to electromagnetic compatibility. It must be installed
and prepared for use as described in chapter 3.
Certain types of mobile telecommunication devices such as mobile telephones are likely to interfere with the M
INI
L.E.D.. The recommended separation distances in this paragraph must therefore be complied with.
The M
INI L.E.D. must not be used near or on top of another device.
If this cannot be avoided, its operation under the conditions of use must be checked beforehand.
The use of accessories other than those specified or sold by SATELEC as replacement parts may have the consequence
of increasing the emissions or decreasing the immunity of the M
INI L.E.D..
10.1 - Electromagnetic emissions
The M
INI L.E.D. is intended for use in the electromagnetic environment specified in the table below. The user and/or
installer must ensure that the M
INI L.E.D. is used in such an environment.
10.2 - Magnetic and electromagnetic immunity
The M
INI L.E.D. ORTHO is intended for use in the electromagnetic environment specified in the table below.
The user and/or installer must ensure that the M
INI L.E.D. ORTHO is used in such an electromagnetic environment.
Emission test Compliance Electromagnetic environment - guidance
RF emission - CISPR
11.
Group 1
Class B
MINI
L.E.D. uses RF energy for internal operation.
Therefore, its radiofrequency emissions are very low and are not likely to cause any
interference in nearby equipment.
MINI L.E.D. uses RF energy for internal operation.
Therefore, its radiofrequency emissions are very low and are not likely to cause any
interference in nearby equipment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Magnetic field at 50Hz.
IEC61000-4-8
3A/m 3A/m
The intensity of the magnetic field should be equivalent to that of
a typical commercial or hospital environment (hospital, clinic).
Electrostatic discharge
(ESD)
IEC 61000-4-2.
± 6KV contact
± 8KV air
± 6KV contact
± 8KV air
Floors must be wood, concrete, cement or tiled.
If floors are covered with synthetic material (carpet, etc.), the
relative humidity must be at least 30%.
Electrical fast transients
IEC 61000-4-4.
± 2KV for power supply
lines
± 2KV for power supply
lines
Power quality should be that of a typical commercial or hospital
environment.
Surges
IEC 61000-4-5.
± 1KV differential mode
± 2KV common mode
± 1KV differential mode
± 2KV common mode
Power quality should be that of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations
IEC 61000-4-11.
<5% UT
T (>95% dip in UT) for
0.5 cycles.
40% UT (60% dip in UT) for 5
cycles
70% UT (30% dip in UT) for
25 cycles
<5% UT (>95% dip in UT) for
250 cycles
<5% U
T (>95% dip in UT) for
0.5 cycles.
40% UT (60% dip in UT) for 5
cycles
70% UT (30% dip in UT) for 25
cycles
<5% UT (>95% dip in UT) for
250 cycles
Power quality should be that of a typical commercial or hospital
environment.
If the use of the MINI L.E.D. requires continued operation during a
power cut, it is recommended that the product be powered from
a separate power supply (UPS, etc.).
ML Classic_S3 V4.qxp 13/03/2008 11:17 Page 9