3
First, read this user manual in its entirety to ensure that the ActivePure® Medical Guardian device is used as intended. Then plug it in, turn it on to the appropriate
fan setting. The ActivePure® Medical Guardian is designed for quiet operation, as not to disturb people in a professional healthcare environment.
BEFORE YOU BEGIN USING THE ACTIVEPURE® MEDICAL GUARDIAN
INTENDED USE
FILL IN AND SAVE
Locate the serial number of your new ActivePure® Medical Guardian and write it here.
Retain it for future reference.
Model No. __________________________________________________________________
Serial No. __________________________________________________________________
Date of Purchase ____________________________________________________________
Distributor _________________________________________________________________
Distributor’s Oce Phone No. __________________________________________________
Distributor’s Oce Address ___________________________________________________
CONTENTS
Important Safety & Use Instructions . . . . . . . . . . . . 2
Intended Use . . . . . . . . . . . . . . . . . . . . . . . 3
Device Description . . . . . . . . . . . . . . . . . . . . 4
Product Features . . . . . . . . . . . . . . . . . . . . . 5
Device Specications . . . . . . . . . . . . . . . . . . . 6
Operation. . . . . . . . . . . . . . . . . . . . . . . . . 6
Treatment Space . . . . . . . . . . . . . . . . . . . . . 7
Settings. . . . . . . . . . . . . . . . . . . . . . . . . . 8
General Maintenance Instructions . . . . . . . . . . . . . 8
Cleaning/Disinfecting and Filter Replacement Instructions . 9
ActivePure® Cell Replacement Instructions. . . . . . .11 - 12
Replacement Parts . . . . . . . . . . . . . . . . . . . .12
FAQs . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Warranty Information . . . . . . . . . . . . . . . . . . .13
Device IEC 60601-1-2:2014 EMC Information . . . . . . . .14
The ActivePure® Medical Guardian, model F170A is a device intended for medical purposes that is used for the reduction of staphylococcus epidermidis and
erwinia herbicola bacteria, MS2 and Phi-X174 viruses and aspergillum niger fungal spores and bacillus globigii bacterial spores from the air in a temperature-
controlled professional healthcare environment of 70~71°F, 40~45% RH.
The ActivePure® Medical Guardian, model F170A has demonstrated the reduction of staphylococcus epidermidis and erwinia herbicola, bacteria MS2 and Phi-X174
viruses and aspergillum niger fungal spores and bacillus globigii bacterial spores under the following conditions;
Organism Type Organism Name Test Temp/RH Exposure Time (m) Avg Log-Reduction
Bacteria Staphylococcus epidermidis 72°F/50% 60 5.95
Bacteria Erwinia herbicola 72°F/50% 60 5.12
Virus MS2 bacteriophage 72°F/50% 60 5.58
Virus Phi-X174 72°F/50% 60 4.19
Fungal spore Aspergillus niger 72°F/50% 60 4.12
Bacterial spore Bacillus globigii 72°F/50% 60 4.22