93-9005-00 ADView 2 User Manual | 31
September 1, 2016
The tested device meets the stated accuracy specification of RMS Error = 1.89% for the range 70% to 100% SaO
2
. The
accuracy is not specifed below 70% SaO
2
. In addition, here are detailed measured A
RMS
values over the specified
discreet ranges:
Range
A
RMS
90-100%
1.54%
80-90%
2.10%
70-80%
2.56%