40
2 liSt autoSeNSe oS
aVerage SCore (%)
2 liSt autoSeNSe oS off
aVerage SCore (%)
Mean (SD) 79.29 (16.439) 54.19 (23.826)
Median 86.98 57.53
Min, Max 45.8, 95.3 19.8, 81.4
In quiet, 100% of subjects’ AutoSense OS scores were within 10% of their AutoSense
OS O scores and in noise 90% of subjects’ AutoSense OS scores were better than their
AutoSense OS O scores by 10% or more. These observed trends further demonstrate
that automatic activation of the AutoSense OS features does not negatively impact
sentence recognition in quiet or noise conditions. Additionally, the improvement
insentence recognition in multi-talker babble noise with AutoSense OS is evident.
Follow-Up Self-Assessment Questionnaire Results
Results were similar across sound quality and listening comfort ratings. All
10subjects (100.0%) reported acceptable sound quality in quiet, with 9 subjects
(90.0%) reporting strong agreement and 1 subject (10.0%) reporting slight
agreement. Nine subjects (90.0%) reported acceptable sound quality in noise
(5subjects reported “Strongly Agree,” 4 subjects reported “Slightly Agree”), with
1subject reporting slight disagreement. All 10 subjects (100.0%) strongly agreed