GUIDANCE AND MANUFACTURER’S DECLARATION
Per IEC 60601-1-2
Electromagnetic Emissions
emiSSioNS teSt ComPliaNCe eleCtromagNetiC eNViroNmeNt - guidaNCe
RF emissions CISPR 11 Group 1 The Naída CI M90 sound processor uses RF
energy only for its internal function. Therefore,
its RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment.
RF emissions CISPR 11 Class B
The Naída CI M90 sound processor is suitable
for use in all establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domestic
purposes. The Naída CI M90 is not intended for
high EM environments such as a MRI scanner
room or near HF surgical electrocautery.
Review the Electromagnetic Immunity section
for determining safe operating environments.
Harmonic emissions
IEC61000-3-2
Not applicable
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Not applicable