Electromagnetic Immunity
The Naída CI M90 sound processor is intended for use in the electromagnetic
environment specified below. The customer or the user of the Naída CI M90 sound
processor should assure that it is used in such an environment.
immuNity ieC 60601
teSt leVel
ComPliaNCe
leVelª
eleCtromagNetiC eNViroNmeNt - guidaNCe
Electrostatic
discharge
(ESD)
IEC61000-4-2
± 8 kV
contact
±15 kV air
± 8 kV
contact
±15 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%. As with the
operation of other electronic devices, precaution
should be taken to not generate ESD.
Power
frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
30 A/m 30 A/m Power frequency magnetic fields should be at
levels characteristic of a typical location in atypical
commercial or hospital environment.