42
Sentences in Noise at 65 dBA, + 5 dB SNR Multi-Talker Babble
Across all subjects, better sentence recognition scores in noise were observed when
using AutoSense OS as compared to the AutoSense OS O program. See table below.
2 liSt autoSeNSe oS
aVerage SCore (%)
2 liSt autoSeNSe oS off
aVerage SCore (%)
N 20 20
Mean (SD) 77.83 (17.397) 48.56 (27.416)
Median 82.62 44.02
Min, Max 31.8, 95.2 9.3, 91.2
For speech recognition in noise, 100% of subjects’ AutoSense OS scores were
improved when compared to their AutoSense OS O scores with an average
improvement of 29.27%. These observed trends demonstrate that automatic activation
of the AutoSense OS features does not negatively impact sentence recognition and
may improve sentence recognition inmulti-talker babble background noise.
In the United States, ClearVoice is approved for pediatric use in children 6 years and
above who 1) are able to complete objective speech perception testing in order to