EasyManua.ls Logo

Advanced Instruments 4D3 - Model 3 D3 Osmometer Specifications

Advanced Instruments 4D3
190 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7
The Advanced
®
3D3/4D3 Service Manual
Model 3D3 Osmometer Specifications
Electrical:
Voltage: 100 to 130 VAC or 200 to 250 VAC
Frequency: 50 or 60 Hz
Fuses: 2-Amp time delay for 100-130 VAC;
1-Amp time delay for 200-250 VAC
Power: 150 Watts
Memory Backup: integral lithium cell;
10 years typical
Sample Volume: 0.20 or 0.25 mL
Sample Capacity: Single sample
Readout: 20-character digital display
Units: mOsm/kg H
2O
Range:
Low: 0 to 2000 mOsm/kg H
2O
High: 1400 to 4000 mOsm/kg H2O
Linearity: Less than ±0.5% from a straight line
Resolution: 1 mOsm/kg H
2O
Repeatability:
0 to 400 mOsm ±2 mOsm/kg (1 S.D.)
400 to 4000 mOsm ±0.5% (1 S.D.)
Communications:
DTE RS-232C, parallel printer, barcode
Storage Temperature: - 40°C to +45°C
(- 40°F to +113°F)
Operating Temperature
(ambient): 18°C to 35°C
(64°F to 95°F)
Room Humidity: 5 to 80% R.H.
(non-condensing)
Drift: Less than 1 digit per month
Ambient Temp. 1 digit for every 5°C (9°F)
Effect: ambient temperature change.
Start-up Time: From stand-by: immediately
From power on: under 5 minutes
Test Time: Approximately 2 minutes per sample
in 0 to 2000 mOsm range; moderately
longer in high range.
Dimensions inches centimeters
Width: 12.5 32
Depth: 16.5 42
Height: 11.5 29
With Head Up: 16 41
Weight pounds kilograms
Net: 27 12
Shipping: 38 18
Warranty: One-year limited warranty on work-
manship and all parts except glass,
plastic and parts warranted by their
makers.
Certification:
Refer to Regulatory Notices (see page 11) for appli-
cable standards.