71P825R Rev1
(4D35 Service Manual)
Page 2 of 2
9. Reset the sample and block probe bin
numbers and the sample stir amplitude
from the service log entries, and any cus-
tomized settings.
10. Reset the serial number (see the following
procedure).
11. Press STOP to exit the Setup menu. The
instrument will run the self-diagnostics
program.
12. Recalibrate the instrument.
Reset Serial Number
1. If there is a Supervisor/Operator
keyswitch, turn it to the Supervisor posi-
tion (if required).
2. At “Press START to Continue”, press
SETUP.
3. Cycle through the menu options by press-
ing < or > until you reach “Serial
Number”, and then press START. The
display will read “Serial#:********”.
4. Press the following keys consecutively:
2, 4, CLEAR. The displayed serial num-
ber will disappear.
5. Enter each digit of the serial number as it
appears on the label on the back of the
instrument (Note: the letter suffix at the
end of the serial number will not be
entered). When the number is correct,
press ENTER.
The serial number will be recorded in
memory and should appear the next time
you power up your instrument.