700P11 Rev0
(4D35 Service Manual)
Page 3 of 3
Fuse Replacement
70011/70022
(Instruments with Serial Suffix P or Greater)
Reference: Use this instruction with replace-
ment part 70011 (1.0-Amp) or
70022 (2.0-Amp).
Voltage/Position/Fuse:
100-130 250V 2.0-Amp time delay
200-250 250V 1.0-Amp time delay
Tools Needed: Flat-bladed screwdriver.
Warning-Hazardous Voltage
CAUTION: Improper connections
may cause damage to the instrument.
Instructions:
1. Turn off the power and unplug the instru-
ment.
2. Use a flat-bladed screwdriver to pry out
the fuse holder door and then the fuse
holder module.
3. Dislodge the old fuses and reinstall the
new fuses by placing them in the fuse
holder brackets.
4. Place the fuse holder back into the instru-
ment, with the correct voltage on the fuse
holder showing in the window (A) on the
fuse holder socket.
If the AC voltage of the power outlet
intended for the instrument is within the
100-130V range, the window should indi-
cate 115V. If the power outlet AC volt-
age is within the 200-250V range, the
window should indicate 230V.
A