EasyManua.ls Logo

Advantage Controls MegaTron - 4-20 mA Interfaces and Calibration; 4-20 mA Output Setup; 4-20 mA Input Setup; Conductivity Calibration

Advantage Controls MegaTron
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
17
2.6 4-20mA Out
Units with a 4-20mA output option will have a menu for setting up the 4-20mA output. The 4mA and 20mA
values can be dened by giving the output proportioning capability. i.e. 4mA = a pH of 6.0 and 20mA = a pH
of 8.0.
SIGNAL SOURCE - Select which probe reading the
mA will use as its reading source.
4 mA VALUE - What the 4mA signal equals
20mA VALUE - What the 20mA signal equals on the
assigned signal sources scale.
2.7 4-20mA Input
SET POINT - What reading turns the relay on
DIFFERENTIAL - Amount reading changes by before
the relay is turned o󰀨
HIGH ALARM - What reading generates a High alarm
notication.
LOW ALARM - What reading generates a Low alarm
notication.
3. Calibration
Calibration is for adjusting the displayed value of a probes reading to match your tester or known solution.
Pick the system or mA input rst. From a particular system pick the probe to calibrate.
CALIBRATION - For adjusting the actual reading
values of the available analog probe inputs, such as
conductivity, pH, ORP or temperature.
3.1 Conductivity Calibration
CALIBRATE - With a clean probe on-line seeing the
system’s water enter the known (tested from a calibrated
hand held tester) value.
RESET ZERO - With the probe out of solution and dry
enter a new zero point. Note: Probe must be dry!
NOTE: In the calibration pop-up window the raw analog to digital (A/D) value will be displayed. A new
calibration value should only be entered when the probe is sensing a stable A/D value. If the system is a
boiler the associated relay can be forced on in the pop-up to allow for a fresh hot sample. Boiler probes
should have fresh hot water blown past them for a minimum of 2 minutes before calibrating.TheA/Dreading
has a range of 0 to 32,767. If it is at one end of the range or the other when trying to calibrate
something is wrong with the probe or wire run.
>OUT 1 SETUP<
SIGNAL SOURCE
4mA VALUE
20mA VALUE
>SYSTEM 1 CALIBRATION<
SYSTEM COND
SYSTEM TEMP
SYSTEM pH
SYSTEM ORP
>SYSTEM 1 COND CALIBRATION<
CALIBRATE Cal Factor 1.10
RESET ZERO
>mA INPUT 1 SETPOINT<
SET POINT
DIFFERENTIAL
HIGH ALARM
LOW ALARM

Other manuals for Advantage Controls MegaTron

Related product manuals