Guidance and manufacture’s declaration electromagnetic immunity
The Wrist Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or
the user of Wrist Blood Pressure Monitor should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
Not applicable
Portable and mobile RF communications
equipment should be used no closer to any part
of the Wrist Blood Pressure Monitor, including
cables, than the recommended separation
distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and is the
recommended separation distance in
meters(m).
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
23