54
CLASSIFICATION OF MEDICAL ELECTRICAL EQUIPMENT
EQUIPMENT CLASSIFICATION
Class IIb Internally Powered Type BF medical device.
CONFORMITY
Directive 93/42/EEC
CellAED® INGRESS PROTECTION
IP 22
MODE OF OPERATION
Continuous, infrequent single event use only
ELECTRODE SPECIFICATIONS
ELECTRICAL CONNECTING CABLE
1 m, 28 AWG
CONTACT AREA
Total combined area: 166 cm
2
ELECTRODE MATERIAL
Hydrogel
ELECTRODE DIMENSIONS
9 cm x 9.25 cm
ENVIRONMENTAL SPECIFICATIONS
OPERATING CONDITIONS
0 ºC to 35 ºC (32 ºF to 95 ºF), 0% to 95% RH (relative humidity) (non-condensing)
STANDBY/STORAGE CONDITIONS
15 ºC to 35 ºC (59 ºF to 95 ºF), 0% to 95% RH (relative humidity) (non-condensing)
TRANSPORT CONDITIONS
0 ºC to 35 ºC (32 ºF to 95 ºF), 0% to 95% RH (relative humidity) (non-condensing)
ALTITUDE
-100 m to 4,000 m
ATMOSPHERIC PRESSURE
60 kPa to 102 kPa