55
SHOCK/DROP/VIBRATION
Transit Operable, Portable Device (EN 60601-1-11:2015, EN 60601-1-6:2010+A1:2015)
ESD/EMI (radiated and immunity)
See electromagnetic conformity tables (Section 10.5, pages 58-61)
UMTS BANDS
800, 850, 900, 1900, 2100
FREQUENCY
GSM 900: 880 MHz ~ 915 MHz, DCS 1800: 1710 MHz ~ 1785 MHz, WCDMA Band 1: 1920 MHz ~ 1980 MHz, , CDMA Band 8:
880 MHz ~ 915 MHz
SAR
GSM 900: 0.192 W/kg, DCS 1800: 0.107 W/kg, WCDMA Band 1: 0.189 W/kg, WCDMA Band 8: 0.059 W/kg
EIRP
GSM900: 31.61 dBm, DCS1800: 32.91 dBm, WCDMA 2100 (Band I): 26.48 dBm, WCDMA 900 (Band VIII): 22.65 dBm
STANDARDS COMPLIANCE
MEDICAL DEVICE
IEC 60601-2-4:2010+AMD1:2018, EN 60601-2-4:2011/A1:2019 IEC 60601-1-11:2015 & EN 60601-1-11:2015
SYMBOLS
ISO 15223-1; UL61010
RESTRICTION OF HAZARDOUS
SUBSTANCE DIRECTIVE (RoHS)
EU Directive 2015/863
ESD/EMI
EN 60601-1-2:2015
EN 60601-2-4:2011+A1:2019
CISPR 11 Group 1 Class B
EN 61000-4-2 Severity Level 4 (Open air discharges up to 8 kV or Direct/Indirect contact discharges up to 6 kV)
BIOCOMPATIBILITY
EN ISO 10993-1:2009/AC:2010
TRANSPORTATION
ANSI ISTA Procedure 3A