MP2C
Manuale operativo - Operating manual MO.MP2C.518
Pag. 60
COMANDI
DI SCRITTURA
WRITING
COMMANDS
MESSAGGIO
MESSAGE
RISPOSTA
ANSWER
VALORE
VALUE
RISOLUZIONE MISURA
MEASUR. RESOLUTION
$RDIINNN<cr> $IIACK<cr> NNN=1 2 5 10 20 50 100
SCARICO
UNLOAD
$ULIIN<cr> $IIACK<cr> N=0=OFF N =1=ON
PUNTO DECIMALE
DECIMAL POINT
$DPIIN<cr> $IIACK<cr>
N = 0÷5
SET-POINT 1
SET-POINT 1
$P1IISNNNNN<cr> $IIACK<cr> NNNNN = 0÷99999
SET-POINT 2
SET-POINT 2
$P2IISNNNNN<cr> $IIACK<cr> NNNNN = 0÷99999
SET-POINT 3
SET-POINT 3
$P3IISNNNNN<cr> $IIACK<cr> NNNNN = 0÷99999
SET-POINT 4
SET-POINT 4
$P4IISNNNNN<cr> $IIACK<cr> NNNNN = 0÷99999
DELTA-POINT 1
HYSTERESIS 1
$D1IINNNN<cr> $IIACK<cr> NNNN = 0÷9999
DELTA-POINT 2
HYSTERESIS 2
$D2IINNNN<cr> $IIACK<cr> NNNN = 0÷9999
DELTA-POINT 3
HYSTERESIS 3
$D3IINNNN<cr> $IIACK<cr> NNNN = 0÷9999
DELTA-POINT 4
HYSTERESIS 4
$D4IINNNN<cr> $IIACK<cr> NNNN = 0÷9999
F.S. DISPLAY
DISPLAY F.S.
$VDIINNNNN<cr> $IIACK<cr> NNNNN = 0÷99999
F.S. USCITA ANALOGICA
ANALOG OUTPUT F.S.
$OAIISNNNNN<cr> $IIACK<cr> NNNNN = 0÷99999
SOPPRESS. della TARA
TARE SUPPRESSION
$CZII<cr> $IIENDCAL<cr>
CAL. PESO CAMPIONE
SAMPLE WEIGHT CAL.
$CPIINNNNN<cr> $IIENDCAL<cr>
NNNNN = 0÷99999
ANNULLAMENTO CAL.
CAL. CANCELLATION
$CRII<cr> $IIACK<cr>
INTESTA. DITTA (campo1)
COMPANY TITLE (field1)
$I1IINNNNNNNNNNNNNNNN<cr> $IIACK<cr> N = ascii char
INTESTA. DITTA (campo2)
COMPANY TITLE (field2)
$I2IINNNNNNNNNNNNNNNN<cr> $IIACK<cr> N = ascii char
INTESTA. DITTA (campo3)
COMPANY TITLE (field3)
$I3IINNNNNNNNNNNNNNNN<cr> $IIACK<cr> N = ascii char