2-98 Subject to Export Control, see Cover Page for details.
2-5-4. MEASURING SWR (cont)
4. With the cursor on the Span Field, press the F1 “Edit” Key and use the Number Keys to
select the desired frequency span. Press the F1 “Done” Key to save the setting.
2
4
6
8
0
1
9
7
5
3
ANT-CA BLE TEST
7.00
200 9-01 -06
11 :35 : 30
SWR
1.00
35.0 MHz
4 5.0
Type : SW R Freq (MHz): 40 .0 Span (MHz): 10.0
Sta rt (MHz): 35.0 Sto p (MHz): 45.0
Edit Return Ho l d Cal Ma r ke r
41 20 29
2
4
6
8
0
1
9
7
5
3
5. Press the F4 “Cal” Key to start the SWR Calibration. Follow the instructions on the screen
to complete the SWR Calibration. When SWR Calibration is completed (“Calibration
Complete” displayed on screen), press the F5 “Done” Key to return to the ANT-Cable Test
Screen.
NOTE: Calibration must be performed at the point the operator is connecting to the
system under test:
If the supplied test cable is being used to connect to the system
under test, calibration is to be performed at the end of the test
cable. Test cable should not exceed four feet in length.
If the system under test cable is being connected directly to the
3500 / 3500A, then calibration is to be performed at the SWR
Connector.
2
4
6
8
0
1
9
7
5
3
ANT-CA BLE TEST
7.00
200 9-01 -06
11 :35 : 30
SWR
1.00
35.0 MHz
4 5.0
Type : SW R Fr eq ( MHz): 40.0 Span (MHz): 10 .0
Sta rt (MHz): 35.0 Sto p (MHz): 45.0
Edit Return Ho l d Cal Ma r ke r
41 20 29
2
4
6
8
0
1
9
7
5
3
NOTE: Once the SWR Connection is calibrated, the SWR Connector remains in the
calibration state until the user changes the point of connection to the UUT. The
ANT-Cable Test Screen displays the Date and Time of the last Calibration above
the Graphical Display.