Subject to Export Control, see Cover Page for details. C-29
C-4. REMOTE OPERATION COMMANDS (cont)
RF Power Meter (cont)
COMMAND RANGE DESCRIPTION
:rfpow:range:dbm:state? 0 - Auto
1 - Manual
2 - Manual - Waiting Update
Returns RF Power Meter autorange state.
:rfpow:range:watts:range? Returns RF Power Meter range information.
:rfpow:range:watts:auto Sets RF Power Meter autorange state to Auto.
:rfpow:range:watts:manual Sets RF Power Meter autorange state to
Manual.
:rfpow:range:watts:state? 0 - Auto
1 - Manual
2 - Manual - Waiting Update
Returns RF Power Meter autorange state.
:rfpow:reading:avg? Returns RF Power Meter average reading.
:rfpow:reading:clear Clears the current minimum, maximum and
average settings.
:rfpow:reading:dbm:avg? 1.0 to 43.0 dBm Returns RF Power Meter average reading.
:rfpow:reading:dbm:max? 1.0 to 43.0 dBm Returns RF Power Meter reading maximum
value.
:rfpow:reading:dbm:min? 1.0 to 43.0 dBm Returns RF Power Meter reading minimum
value.
:rfpow:reading:dbm:val? 1.0 to 43.0 dBm Returns RF Power Meter average reading.
:rfpow:reading:max? 1.0 to 43.0 dBm Returns RF Power Meter reading maximum
value.
:rfpow:reading:min? 1.0 to 43.0 dBm Returns RF Power Meter reading minimum
value.
:rfpow:reading:val? 1.0 to 43.0 dBm Returns RF Power Meter average reading.
:rfpow:reading:watt:avg? 0.00125 to 20 W Returns RF Power Meter average reading.
:rfpow:reading:watt:max? 0.00125 to 20 W Returns RF Power Meter reading maximum
value.
:rfpow:reading:watt:min? 0.00125 to 20 W Returns RF Power Meter reading minimum
value.
:rfpow:reading:watt:val? 0.00125 to 20 W Returns RF Power Meter average reading.
:rfpow:state Enables/disables RF Power Meter operation.
:rfpow:state? Returns state of RF Power Meter.
:rfpow:units 0 - dBm
1 - Watts
Sets units for reading.
:rfpow:zero Starts zero operation required before
measurements.