AgaMatrix Presto - Troubleshooting Out-of-Range Control Results; Common Causes for Out-of-Range Results

AgaMatrix Presto
62 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
21
from the control solution tests continue to fall outside that
range, do not use the system to test your blood glucose. Call
customer service.
IMPORTANT: Out-of-range results may be due to one or
more of the following factors:
[1] Expired or defective control solution.
[2] Expired or defective test strip.
[3] Error in performing test.
[4] Watered-down control solution.
[5] Meter malfunction.
[6] Control solution test done outside the system operating
temperature of 50 °F to 104 °F (10 °C to 40 °C).
[7] Failure to shake the control solution bottle vigorously
before using.
[8] Failure to discard the  rst drop of control solution and to
wipe the bottle tip clean.