DX- D 600
User Manual
0242F EN 20181218
144
Table 8-1 (Cont.)
System/Miscellaneous
CODE USER ERROR MESSAGE ERROR DESCRIPTIO N ACTION
20 Initial position unreachable Sequence Initial position unreachable Correct and try again
21
Invalid parameter range: Unreachable
positions
Sequence Invalid parameter range:
Unreachable positions
Correct and try again
22 FLFS Sequence Aborted Stitching Error Try again
23 FLFS Sequence Completed Stitching Error Try again
24 FLFS: Collimator Key in Manual Mode Stitching error Correct and try again
25 FLFS: Detector NOT in Portrait Mode Stitching Error Try again
26
FLFS: Incorrect Parameters: Check
Grid, Top and Bottom
Stitching Error Check Grid, Top and Bottom
27 FLFS: Calculating Geometry Stitching Error Try again
28 Stitching: Calculating Geometry Stitching Error Try again
29
Stitching: Detector NOT in Portrait
Mode
Stitching Error Correct and try again
11001 Read Error in Calibration File
File not present or opened
by other application
Try again
11002 W rite Error in Calibration File File not present or locked Try again
11003 Position Demand Error
Not possible to reach to the configured
position
Modify position configuration
11004 Workstation not available Invalid Workstation Configure correct Workstation
11005
Tomography: Parameters not
Configured
Tomography not configured yet Configure Tomography parameters
11006
Emergency OFF Switch pressed:
Synchronizing Positioner software
resources: Please wait
Emergency OFF Switch pressed
Release Emergency OFF Switch after
the emergency cause disappears
11007
Safety issue:
Keep OUT: Do not operate
Safety Error Do not operate. Contact Service
11008
CS/Wall Stand: Safety area around
the Wall Stand Bucky/Detector
crossed
For automatic movements, move out
of the safety area manually
11009
CS/Table: Safety area around
the Table Bucky/Detector is crossed
For automatic movements, move out
of the safety area manually
11010
Operative Rejected: The
Auto-movement interlock is activated:
Please, release signal state to
proceed
Try again