Technical and Performance Specifications
Ta bl e 6 Performance specifications
Type Comment
High breaking capacity fuse 2.0 A, 250 V
Signal specifications:
CAN 2-wire serial bus system
Analog-input Input Range= -1 V to 1 V
Resolution = 24 bit A/D-Converter
Effective System Resolution = +/- 1 uV (rms)
GPIO TTL compatible Input and Output
Open collector characteristic Input voltage range 0 V to 5 V
Relay outputs 2 contact closure outputs and 2 double throw outputs
I
max
= 250 mA
U
max
= 30 V DC/AC
The contacts are protected by a self-resetting PTC-fuse