Calibration Error Messages
Error Error Message
701 Cal security disabled by jumper
702 Cal secured
703 Invalid secure code
704 Secure code too long
705 Cal aborted
706 Cal value out of range
707 Cal signal measurement out of range
708 Flatness cal failed
709 Cannot calibrate frequency while externally locked (Option 001)
760 RAM checksum failure
850 Cal setup invalid
851 Negative offset gain cal required (CAL:SETup 50)
852 Flatness DAC gain cal required (CAL:SETup 64)
853 AM cal 1 required (CAL:SETup 30)
854 AM cal 2 required (CAL:SETup 31)
855 Cal load resistance not specified (CAL:SETup 33)
856 Square wave positive offset cal required (CAL:SETup 60)
857 Square wave 50% duty cycle cal required (CAL:SETup 62)
858 AM cal 3 required (CAL:SETup 32)
Chapter 4 Calibration Procedures
Error Messages
82