4349B
SYST
em
Subsystem
SYSTem
Subsystem
The
SYST
em
subsystem
reports the
rmware version
and errors
,sets
the beeper
, locks
out
front-panel
key
,
and
denes the
power LINE
frequency.
COMMAND
PARAMETER
SYST
em
:BEEP
er
[:IMMediate]
:ST
AT
e
<
Boolean
>
:ERRor?
:KLOCk
<
Boolean
>
:LFRequency
<
numeric
value
>
:PRESet
:VERSion?
:SYST
em:BEEP
er[:IMMediate]
Emits
a
beep
.
(no
query)
:SYST
em:BEEP
er:ST
A
T
e
f
ON
j
OFF
j
1
j
0
g
Sets
or
queries if
the beeper
is
enabled.
ON
,
1
enables
the beeper
output,
and
set
beeper
mode
as
dened
by
CALCulate{1|2|3|4}:LIMit:BEEPer
command.
OFF
,
0
disables
the all
beeper
output,
including
error
beep
.
Query
response
is
{1|0}
.
:SYST
em:ERRor?
Returns
the
existing
error
numbers
and
messages
in
the
4349B
's
error
queue
.
(Query
only)
:SYST
em:KLOCk
f
ON
j
OFF
j
1
j
0
g
Sets
or
queries
whether
or not
the front-panel
keys of
the
4349B
locked.
ON
,
1
the
front-panel keys
are locked.
OFF
,
0
the
front-panel keys
are unlocked.
Query response is
{1|0}
.
:SYSTem:LFRequency
<
numeric
value
>
Sets or queries the 4349B 's operating power line frequency
.A
vailable line frequency selections
are 50 Hz and 60 Hz.
<
numeric value
>
is,
numeric
50 or 60 (integer)
GPIB Reference 5-27

Table of Contents

Other manuals for Agilent Technologies 339B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 339B and is the answer not in the manual?

Agilent Technologies 339B Specifications

General IconGeneral
BrandAgilent Technologies
Model339B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals