EasyManuals Logo

Agilent Technologies 54621A Service Guide

Agilent Technologies 54621A
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #64 background imageLoading...
Page #64 background image
3-20
Agilent 54622A/22D/24A Performance Test Record
Agilent 54622A/54622D/54624A
Serial No. ______________________________________ Test by _____________________________
Test Interval ____________________________________ Work Order No. ______________________
Recommended Next Testing ________________________ Temperature ____________
Threshold Specification Limits Ch D7-D0 Ch D15-D8
Accuracy Test
5 V - 250 mV 4.750 V ________ ________
(100 mV + 3% of
threshold setting)
5 V + 250 mV 5.250 V ________ ________
-5 V - 250 mV -5.250 V ________ ________
-5 V + 250 mV -4.750 V ________ ________
0 V - 100 mV -100 mV ________ ________
0 V + 100 mV 100 mV ________ ________
Voltage Measurement Accuracy
Range Power Supply Setting Test Limits Channel 1 Channel 2 Channel 3 Channel 4
5 V/Div 35 V 34.04 V to 35.96 V ________ ________ ________ ________
2 V/Div 14 V 13.616 V to 14.384 V ________ ________ ________ ________
1 V/Div 7 V 6.808 V to 7.192 V ________ ________ ________ ________
500 mV/Div 3.5 V 3.404 V to 3.596 V ________ ________ ________ ________
200 mV/Div 1.4 V 1.3616 V to 1.4384 V ________ ________ ________ ________
100 mV/Div 700 mV 680.8 mV to 719.2 mV ________ ________ ________ ________
50 mV/Div 350 mV 340.4 mV to 359.6 mV ________ ________ ________ ________
20 mV/Div 140 mV 136.16 mV to 143.84 mV ________ ________ ________ ________
10 mV/Div 70 mV 68.08 mV to 71.92 mV ________ ________ ________ ________
5 mV/Div 35 mV 34.04 mV to 35.96 mV ________ ________ ________ ________
2 mV/Div 14 mV 13.616 mV to 14.384 mV ________ ________ ________ ________
1 mV/Div 7 mV 6.616 mV to 7.384 mV ________ ________ ________ ________
Bandwidth Test Limits Channel 1 Channel 2 Channel 3 Channel 4
3 dB at 100 MHz ________ ________ ________ ________
Horizontal t and 1/t Accuracy
Generator Setting Test Limits Results
Frequency 10 kHz 9.98 kHz to 10.02 kHz ________
Period 100
µs 99.79 µs to 100.2 µs ________
Frequency 1 MHz 997.9 kHz to 1.002 MHz ________
Period 1
µs 997.9 ns to 1.002 µs ________
Frequency 100 MHz 99.10 MHz to 100.9 MHz ________
Period 10 ns 9.91 ns to 10.09 ns ________
Trigger Sensitivity Test Limits Channel 1 Channel 2 Channel 3 Channel 4
Internal trigger 25 MHz at 0.35 division ________ ________ ________ ________
100 MHz at 0.35 division ________ ________ ________ _______
Ext
External trigger 25 MHz at <75 mV
p-p
________
100 MHz at <150 mV
p-p
________
service.book Page 20 Wednesday, December 18, 2002 8:35 AM

Other manuals for Agilent Technologies 54621A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 54621A and is the answer not in the manual?

Agilent Technologies 54621A Specifications

General IconGeneral
BrandAgilent Technologies
Model54621A
CategoryTest Equipment
LanguageEnglish

Related product manuals