EasyManua.ls Logo

Agilent Technologies 54641A - Page 57

Agilent Technologies 54641A
124 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-13
Testing Performance
To verify bandwidth
To verify bandwidth
This test verifies bandwidth. In this test you will use an oscilloscope calibrator
with a level sinewave output.
You will use the peak-to-peak voltage both at 1 MHz and at bandwidth frequency
to verify the bandwidth response of the oscilloscope.
54642A and 54642D
Test limits at 2 mV/div to 5 V/div:
All channels (±3 dB)
dc to 500 MHz
ac coupled 3.5 Hz to 500 MHz
54641A and 54641D
Test limits at 2 mV/div to 5 V/div:
All channels (±3 dB)
dc to 350 MHz
ac coupled 3.5 Hz to 350 MHz
Table 3-6
Equipment Required to Verify Bandwidth
Equipment Critical Specifications Recommended Model/Part
Oscilloscope Calibrator Fluke 5820A
Cable * Type N (m), 24-inch Agilent 11500B
* The oscilloscope calibrator is supplied with 2 or more coaxial cables N (m), BNC (m),
1 meter long, Fluke P/N 686318.
service.book Page 13 Wednesday, August 28, 2002 3:25 PM

Related product manuals