EasyManua.ls Logo

Agilent Technologies 54641A - Page 71

Agilent Technologies 54641A
124 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-5
Calibrating and Adjusting
To adjust the power supply
2
Locate the power supply voltages at L200, L201, L202, and L203 on the
system board.
The power supply voltages are not labeled on the system board. See
Figure 4-1 for the locations.
Figure 4-1
Low Voltage Power Supply Voltages (on the bottom of the oscilloscope)
3 Make sure that the voltage measurements are within the tolerances
listed in Table 4-2.
Table 4-2 Power Supply Voltage Tolerances
E
-
- -
9
/
9
9
9
-
-
/
/
/
Supply Voltage Tolerance
+5.1 V ±250 mV (4.85 V to 5.35 V)
-5.2 V ±156 mV (-5.04 V to -5.36 V)
+15.75 V +1.260 V, -787 mV (+14.963 V to +17.010 V)
+3.4 V ±100 mV (+3.30 V to +3.50 V)
+8.2 V ±82 mV (+8.12 V to +8.28 V)
measure at Printer Power connector on rear panel
of instrument
service.book Page 5 Wednesday, August 28, 2002 3:25 PM

Related product manuals