41
Table 1-6. Supplemental GPIB Characteristics for All Models
-
Parameter
All Models
Digital Port Characteristics
Maximum ratings:
16.5Vdc between terminals 1&2; 3&4; and from 1 or 2 to chassis
ground
FLT/INH Operation
FLT/INH Terminals 1 & 2:
I
ol
(low-level output current)
V
ol
(low-level output voltage)
1.25 mA maximum
0.5 V maximum
FLT/INH Terminals 3 & 4:
V
il
(low-level input voltage)
V
ih
(high-level input voltage)
I
il
(low-level input current)
tw (pulse width)
td (time delay)
0.8 V maximum
2.0 V minimum
1 mA
100 μs, minimum
4 ms, typical
Digital I/O Operation
Digital OUT Port 0,1,2 - Open Collector:
I
oh
(high-level output leakage @ 16.5 V)
I
oh
(high-level output leakage @ 5.25 V)
I
ol
(low-level output sink current @ 0.5 V)
I
ol
(low-level output sink current @ l V)
100 μA (ports 0,1); 12.5 mA (port 2)
100 μA (ports 0,1); 250 μA (port 2)
4 mA
250 mA
Digital IN Port 2 - Internal 4.64 k Pullup:
I
il
(low-level input current @ 0.4 V)
I
ih
(high-level input current @ 5.25 V)
V
il
(low-level input voltage)
V
ih
(high-level input voltage)
1.25 mA
250 μA
0.8 V maximum
2.0 V minimum
GPIB Interface Capabilities
Languages:
Interface:
SCPI (default); Compatibility
AH1, C0, DC1, DT1, E1, LE4, PP0, RL1, SH1, SR1, TE6
Serial Link Capabilities (multiple supplies sharing one GPIB primary address)
Maximum number of supplies:
Maximum number of linked supplies:
Maximum total chain cable length:
16
15
30 m (100 ft)
Table 1-7. Operator Replaceable Parts List
Description
(Unless otherwise specified, parts apply to all models.)
Agilent Part No.
Cable assembly, GPIB
Cable assembly, serial link
Collar, rotary output control
Cover, ac input safety
Series 667xA, w/strain relief connector & rubber boot
Series 668xA, w/strain relief connector & rubber boot
Series 669xA, w/strain relief connector & rubber boot
Cover, dc output
Series 664xA and 665xA
Series 667xA
Series 668xA
Series 669xA
Flatwasher, ac input safety cover (Series 667xA, 668xA)
Foot, cabinet
(See “Accessories”)
(See “Accessories”)
5040-l700
5040-1676
5060-3237
5065-6933
0360-2191
5040-1674
5040-1692
5040-1692
3050-1053
5041-8801

Table of Contents

Other manuals for Agilent Technologies 6680A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6680A and is the answer not in the manual?

Agilent Technologies 6680A Specifications

General IconGeneral
BrandAgilent Technologies
Model6680A
CategoryPower Supply
LanguageEnglish

Related product manuals