SERVICE MANUAL
GPIB DC Power Supplies
Series 657xA and 667xA
For instruments with Serial Numbers:
Agilent Model 6571A: US36520101 and Above *
Agilent Model 6572A: US36350101 and Above *
Agilent Model 6573A: US36330101 and Above *
Agilent Model 6574A: US36340101 and Above *
Agilent Model 6575A: US36340101 and Above *
Agilent Model 6671A: US36400101 and Above *
Agilent Model 6672A: US36390101 and Above *
Agilent Model 6673A: US36380101 and Above *
Agilent Model 6674A: US36370101 and Above *
Agilent Model 6675A: US36370101 and Above *
* For instruments with higher serial numbers, a change page may be included.
For instruments with lower serial numbers, see Appendix A.
Agilent Part No. 5961-2583 Printed in USA
Microfiche Part No. 5961-2584 September 2000

Table of Contents

Questions and Answers

  • J
    Jonathan RuizAug 6, 2025
    Why did the front panel EEPROM checksum test fail on my Agilent Technologies 6675A Power Supply?
    • L
      Louis SimpsonAug 6, 2025
      The front panel EEPROM checksum test might have failed on your Agilent Technologies Power Supply possibly due to power loss during a write operation. If power loss is not the problem, the EEPROM A1U6 could be defective.
  • C
    Christine GuerreroAug 11, 2025
    What causes EEPROM write failure in Agilent Technologies Power Supply?
    • D
      donald30Aug 11, 2025
      EEPROM write failure in Agilent Technologies Power Supply may be caused by a defective EEPROM A1U6 or a calibration error.
  • S
    Sharon MillerAug 16, 2025
    What causes the front panel ROM test to fail on Agilent Technologies 6675A Power Supply?
    • J
      Jeffrey PriceAug 17, 2025
      The front panel ROM test failure on Agilent Technologies Power Supply may be caused by ROM AlU4 or address latches AlU8.
  • M
    Megan JenkinsAug 21, 2025
    What causes CV or CC DAC tests to fail in Agilent Technologies 6675A?
    • W
      weavertylerAug 21, 2025
      CV or CC DAC tests failure in Agilent Technologies Power Supply may be caused by CV DAC A5U507/U508 or CC DAC A5U509/U510.
  • D
    David MillerAug 27, 2025
    What could be the reason for ambient temperature readback test failure in Agilent Technologies Power Supply?
    • M
      Megan BrownAug 27, 2025
      The ambient temperature readback test failure in Agilent Technologies Power Supply may be caused by thermistor A5RT770 or comparator.
  • T
    Tara YoungAug 31, 2025
    What could be the reason for secondary interface 5 volt readback test failure in Agilent Technologies 6675A?
    • J
      Jessica CohenAug 31, 2025
      Secondary interface 5 volt readback test failure in Agilent Technologies Power Supply may be caused by comparators A5U513 or a failed readback DAC.

Agilent Technologies 6675A Specifications

General IconGeneral
Model6675A
TypeDC Power Supply
Output Current0 to 5 A
Number of Outputs1
Programming Accuracy Voltage0.05% + 10 mV
Programming Accuracy Current0.1% + 5 mA
Input Voltage100 V / 120 V / 220 V ±10% or 240 V +5% -10%
Voltage Programming Accuracy0.05% + 10 mV
Current Programming Accuracy0.1% + 5 mA
Operating Temperature Range0 to 50 °C
Storage Temperature Range-40 to 70 °C
Input Frequency47 to 63 Hz
Load Regulation0.01% + 2mV
Line Regulation0.01% + 2mV
Ripple and Noise2 mV RMS
Ripple & Noise (Voltage)2mV rms, 10mV p-p
Voltage Ripple and Noise (20 Hz to 20 MHz)2mV rms, 10mV p-p
Relative HumidityUp to 95% RH (non-condensing)
Efficiency70% typical