SERVICE MANUAL
GPIB DC Power Supplies
Series 657xA and 667xA
For instruments with Serial Numbers:
Agilent Model 6571A: US36520101 and Above *
Agilent Model 6572A: US36350101 and Above *
Agilent Model 6573A: US36330101 and Above *
Agilent Model 6574A: US36340101 and Above *
Agilent Model 6575A: US36340101 and Above *
Agilent Model 6671A: US36400101 and Above *
Agilent Model 6672A: US36390101 and Above *
Agilent Model 6673A: US36380101 and Above *
Agilent Model 6674A: US36370101 and Above *
Agilent Model 6675A: US36370101 and Above *
* For instruments with higher serial numbers, a change page may be included.
For instruments with lower serial numbers, see Appendix A.
Agilent Part No. 5961-2583 Printed in USA
Microfiche Part No. 5961-2584 September 2000

Table of Contents

Questions and Answers

  • C
    Courtney SmithAug 6, 2025
    Why did the front panel EEPROM checksum test fail on my Agilent Technologies 6674A?
    • M
      mary10Aug 6, 2025
      The front panel EEPROM checksum test might have failed on your Agilent Technologies Power Supply possibly due to power loss during a write operation. If power loss is not the problem, the EEPROM A1U6 could be defective.
  • W
    Whitney JenkinsAug 11, 2025
    What causes EEPROM write failure in Agilent Technologies 6674A?
    • R
      Ryan HicksAug 11, 2025
      EEPROM write failure in Agilent Technologies Power Supply may be caused by a defective EEPROM A1U6 or a calibration error.
  • A
    Angel JonesAug 16, 2025
    What causes the front panel ROM test to fail on Agilent Technologies 6674A Power Supply?
    • K
      Kristin ChambersAug 16, 2025
      The front panel ROM test failure on Agilent Technologies Power Supply may be caused by ROM AlU4 or address latches AlU8.
  • C
    Christian OconnorAug 21, 2025
    What causes CV or CC DAC tests to fail in Agilent Technologies 6674A Power Supply?
    • A
      Aaron BaldwinAug 21, 2025
      CV or CC DAC tests failure in Agilent Technologies Power Supply may be caused by CV DAC A5U507/U508 or CC DAC A5U509/U510.
  • M
    mike52Aug 27, 2025
    What could be the reason for ambient temperature readback test failure in Agilent Technologies Power Supply?
    • M
      Mary FlynnAug 27, 2025
      The ambient temperature readback test failure in Agilent Technologies Power Supply may be caused by thermistor A5RT770 or comparator.
  • J
    Janet AndersonAug 31, 2025
    What could be the reason for secondary interface 5 volt readback test failure in Agilent Technologies Power Supply?
    • M
      mark50Aug 31, 2025
      Secondary interface 5 volt readback test failure in Agilent Technologies Power Supply may be caused by comparators A5U513 or a failed readback DAC.

Agilent Technologies 6674A Specifications

General IconGeneral
Output Voltage0 to 60 V
Number of Outputs1
Line Regulation0.01% + 2 mV
Programming Accuracy Current0.1% + 5 mA
Input Frequency47 to 63 Hz
Operating Temperature Range0 to 50 °C
Storage Temperature Range-40 to 70 °C
CoolingForced air
Current Ripple and Noise (20Hz to 20MHz)5 mArms
Relative Humidity0 to 95% non-condensing
Load Regulation0.01% + 2 mV